Abstract
ABSTRACT
The increasing demand for molecular diagnostics in clinical microbiology laboratories necessitates automated sample processing. In the present study, we evaluated the performance of the MagNA Pure LC total nucleic acid isolation kit (M extraction) in comparison with the manual method (Si extraction) according to Boom et al. (R. Boom, C. J. A. Sol, M. M. M. Salimans, C. L. Jansen, P. M. Wertheim-van Dillen, and J. van der Noordaa, J. Clin. Microbiol. 28:495-503, 1990) for the detection of viral DNA by competitive quantitative PCR. Reconstruction experiments with HindIII-digested phage lambda DNA and HaeIII-digested φX174 DNA showed that the recovery of DNA from phosphate-buffered saline, cerebrospinal fluid, EDTA-anticoagulated plasma, and EDTA-anticoagulated whole blood by M extraction is, on average, 6.6-fold lower compared to Si extraction. PCR signals of spiked PCR control DNAs for Epstein-Barr virus and varicella-zoster virus were also between 1.9- and 14.2-fold lower after M extraction compared to Si extraction, also suggesting impaired DNA recovery. M extraction of spiked cytomegalovirus strain AD 169 in whole blood showed a 5- to 10-fold reduction in PCR sensitivity compared to Si extraction. This reduction of PCR sensitivity was also observed when clinical whole blood samples were processed by M extraction. Before implementing M extraction, the clinical consequences of the reduced recovery should first be considered, especially when maximal sensitivity is required.
Publisher
American Society for Microbiology
Cited by
37 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献