Discordance between Viral Loads Determined by Roche COBAS AMPLICOR Human Immunodeficiency Virus Type 1 Monitor (Version 1.5) Standard and Ultrasensitive Assays Caused by Freezing Patient Plasma in Centrifuged Becton-Dickinson Vacutainer Brand Plasma Preparation Tubes

Author:

Salimnia Hossein1,Moore Ellen C.2,Crane Lawrence R.3,MacArthur Rodger D.3,Fairfax Marilynn R.1

Affiliation:

1. Department of Pathology and DMC University Laboratories

2. Division of Allergy, Immunology and Rheumatology, Department of Pediatrics

3. Division of Infectious Diseases, Department of Medicine, Wayne State University School of Medicine, 540 E. Canfield, Detroit, Michigan 48201

Abstract

ABSTRACT The Roche COBAS AMPLICOR human immunodeficiency virus type 1 (HIV-1) Monitor (version 1.5) standard and ultrasensitive viral load assays often gave discordant results, with viral loads from the standard assay exceeding those from the ultrasensitive assay by more than 0.5 log 10 for approximately 20% of specimens received. We began studies to determine the extent, magnitude, and reproducibility of the discordance between the assays and to discover and eliminate the cause of this discordance. Until then, we revised our standard operating procedure to include both standard and ultrasensitive testing on all specimens submitted for viral load determinations. Discordant results usually recurred on retesting. They were most prevalent for specimens with ultrasensitive viral loads of <1,000 and rare for specimens with viral loads of >10,000. Often, standard assay results exceeded those of the ultrasensitive assay by 50- to 100-fold. At higher viral loads, the difference between the standard and ultrasensitive assays persisted, but the percent difference was smaller and rarely caused discordance. The proportion of discordant results was significantly higher in specimens from pediatric patients than in specimens from adults. The ultrasensitive viral load determinations generally agreed with the results of the B-DNA (Bayer) viral load assays. If the plasma was transferred from the centrifuged plasma preparation tubes before freezing, standard and ultrasensitive results were concordant with each other and with values determined on plasma from lavender-topped EDTA tubes.

Publisher

American Society for Microbiology

Subject

Microbiology (medical)

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3