Affiliation:
1. Henry Ford Health System, Department of Pathology, Detroit, Michigan, USA
Abstract
ABSTRACT
Rapid identification of pathogens directly from positive blood cultures can play a major role in reducing patient mortality rates. We evaluated the performance of the Verigene Gram-Positive Blood Culture (BC-GP) assay (Nanosphere Inc., Northbrook, IL) for detection of commonly isolated Gram-positive organisms as well as associated resistance markers from positive blood cultures. Positive blood cultures (VersaTREK; Trek Diagnostic Systems, Independence, OH) from 203 patients with Gram-positive organism infections were analyzed using the BC-GP assay within 12 h for the detection of 12 different organisms, including staphylococci, streptococci, and enterococci, as well as for the presence of 3 resistance markers (
mecA
,
vanA
, and
vanB
). Results were compared to those of routine laboratory methods for identification and susceptibility testing. For identification of organisms and detection of resistance markers in 178 monomicrobial positive blood cultures, the BC-GP assay showed 94% and 97% concordance, respectively, with routine methods. After 25 polymicrobial cultures were included, the results showed 92% and 96% agreement for identification and resistance markers, respectively, for a total of 203 positive cultures. In 6/25 polymicrobial cultures, at least 1 isolate was not detected. Concordance levels for detection of major pathogens such
Staphylococcus aureus
(
n
= 45) and enterococci (
n
= 19) were 98% and 95%, respectively. Agreement levels for detection of resistance markers such as
mecA
and
vanA
/
B
were 92% and 100%, respectively. The BC-GP assay is capable of providing rapid identification of Gram-positive cocci as well as detection of resistance markers directly from positive blood cultures at least 24 to 48 h earlier than conventional methods.
Publisher
American Society for Microbiology
Cited by
85 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献