Affiliation:
1. Division of Clinical Microbiology, Department of Laboratory Medicine and Pathology, Mayo Clinic, Rochester, Minnesota, USA
Abstract
ABSTRACT
Detection of
Cryptococcus
antigen (CrAg) is invaluable for establishing cryptococcal disease. Multiple different methods for CrAg detection are available, including a lateral flow assay (LFA). Despite excellent performance of the CrAg LFA, we have observed multiple cases of low-titer (≤1:5) positive CrAg LFA results in patients for whom cryptococcosis was ultimately excluded. To investigate the accuracy of low-titer positive CrAg LFA results, we performed chart reviews for all patients with positive CrAg LFA results between June 2014 and December 2016. During this period, serum and/or cerebrospinal fluid (CSF) samples from 3,969 patients were tested with the CrAg LFA, and 55 patients (1.5%) tested positive. Thirty-eight of those patients lacked a history of cryptococcal disease and were the focus of this study. Fungal culture or histopathology confirmed
Cryptococcus
infection for 20 patients (52.6%), and CrAg LFA titers in serum and CSF samples ranged from 1:5 to ≥1:2,560. For the 18 patients (47.4%) without culture or histopathological confirmation, the CrAg LFA results were considered true-positive results for 5 patients (titer range, 1:10 to ≥1:2,560), due to clinical improvement with targeted therapy and decreasing CrAg LFA titers. The remaining 13 patients had CrAg LFA titers of 1:2 (
n
= 11) or 1:5 (
n
= 2) and were ultimately diagnosed with an alternative condition (
n
= 11) or began therapy for possible cryptococcosis without improvement (
n
= 2), leading to an overall CrAg LFA false-positive rate of 34%. We recommend careful clinical correlation prior to establishing a diagnosis of cryptococcal infection for patients with first-time positive CrAg LFA titers of 1:2.
Publisher
American Society for Microbiology
Cited by
40 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献