Accuracy of Commercial and Reference Susceptibility Testing Methods for Detecting Vancomycin-Intermediate Staphylococcus aureus

Author:

Swenson Jana M.1,Anderson Karen F.1,Lonsway David R.1,Thompson Angela1,McAllister Sigrid K.1,Limbago Brandi M.1,Carey Roberta B.1,Tenover Fred C.1,Patel Jean B.1

Affiliation:

1. Clinical and Environmental Microbiology Branch, Division of Healthcare Quality Promotion, Centers for Disease Control and Prevention, Atlanta, Georgia 30333

Abstract

ABSTRACT We compared the results obtained with six commercial MIC test systems (Etest, MicroScan, Phoenix, Sensititre, Vitek Legacy, and Vitek 2 systems) and three reference methods (agar dilution, disk diffusion, and vancomycin [VA] agar screen [VScr]) with the results obtained by the Clinical and Laboratory Standards Institute broth microdilution (BMD) reference method for the detection of VA-intermediate Staphylococcus aureus (VISA). A total of 129 S. aureus isolates (VA MICs by previous BMD tests, ≤1 μg/ml [ n = 60 strains], 2 μg/ml [ n = 24], 4 μg/ml [ n = 36], or 8 μg/ml [ n = 9]) were selected from the Centers for Disease Control and Prevention strain collection. The results of BMD with Difco Mueller-Hinton broth were used as the standard for data analysis. Essential agreement (percent ±1 dilution) ranged from 98 to 100% for all methods except the method with the Vitek Legacy system, for which it was 90.6%. Of the six commercial MIC systems tested, the Sensititre, Vitek Legacy, and Vitek 2 systems tended to categorize VISA strains as susceptible (i.e., they undercalled resistance); the MicroScan and Phoenix systems and Etest tended to categorize susceptible strains as VISA; and the Vitek Legacy system tended to categorize VISA strains as resistant (i.e., it overcalled resistance). Disk diffusion categorized all VISA strains as susceptible. No susceptible strains (MICs ≤ 2 μg/ml) grew on the VScr, but all strains for which the VA MICs were 8 μg/ml grew on the VScr. Only 12 (33.3%) strains for which the VA MICs were 4 μg/ml grew on VScr. The differentiation of isolates for which the VA MICs were 2 or 4 μg/ml was difficult for most systems and methods, including the reference methods.

Publisher

American Society for Microbiology

Subject

Microbiology (medical)

Reference10 articles.

1. Appelbaum, P. C. 2007. Reduced glycopeptide susceptibility in methicillin-resistant Staphylococcus aureus (MRSA). Int. J. Antimicrob. Agents30:398-408.

2. CLSI document M7-A7. 2006

3. CLSI document M2-A9. 2006

4. CLSI document M100-S19 2009

5. Jones, R., M. Stilwell, H. Sader, T. Fritsche, and B. P. Goldstein. 2006. Spectrum and potency of dalbavancin tested against 3322 gram-positive cocci isolated in the United States Surveillance Program (2004). Diagn. Microbiol. Infect. Dis.53:149-153.