Affiliation:
1. The Lilly Research Laboratories, Eli Lilly & Co., Indianapolis, Indiana 46206
Abstract
Several factors are investigated that normally cause variation in zone diameters in conventional disc plate diffusion assay procedures. Of these factors the most serious is the unequal exposure of the individual plates at top or bottom of stacks to temperatures above and below room temperature. This unequal temperature exposure is avoided by novel handling and incubation procedures. A major variable, but one which can be controlled, is the varying time interval between pouring seeded agar and the time of applying the pads with antibiotic to the plates. This influence of time of setting and the effects of several other sequential operations are combined into a composite variable. This variable is then accounted for and normalized by interposing “external” reference plates set with a reference solution in the sequence of approximately 100 plates. No “internal” reference zones are employed. Such factors as volume of agar poured, wedge shape of agar in a dish, volumetric errors in dilutions, and timing considerations are studied and discussed. The results of this study form the basis for a test protocol which is presented in a following paper.
Publisher
American Society for Microbiology
Subject
General Pharmacology, Toxicology and Pharmaceutics,General Immunology and Microbiology,General Biochemistry, Genetics and Molecular Biology,General Medicine
Cited by
92 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献