Performance of Three Commercial Viral Load Assays, Versant Human Immunodeficiency Virus Type 1 (HIV-1) RNA bDNA v3.0, Cobas AmpliPrep/Cobas TaqMan HIV-1, and NucliSens HIV-1 EasyQ v1.2, Testing HIV-1 Non-B Subtypes and Recombinant Variants

Author:

Holguín Africa12,López Marisa1,Molinero Mar1,Soriano Vincent1

Affiliation:

1. Department of Infectious Diseases, Hospital Carlos III, Madrid, Spain

2. HIV-1 Molecular Epidemiology Laboratory, Department of Microbiology, Hospital Universitario Ramón y Cajal and CIBER-ESP, Madrid, Spain

Abstract

ABSTRACT Monitoring antiretroviral therapy requires that human immunodeficiency virus type 1 (HIV-1) viremia assays are applicable to all distinct variants. This study evaluates the performance of three commercial viral load assays—Versant HIV-1 RNA bDNA v3.0, Cobas AmpliPrep/Cobas TaqMan HIV-1, and NucliSens HIV-1 EasyQ v1.2—in testing 83 plasma specimens from patients carrying HIV-1 non-B subtypes and recombinants previously defined by phylogenetic analysis of the pol gene. All 28 specimens from patients under treatment presented viremia values below the detection limit with the three methods. In the remaining 55 specimens from naive individuals viremia could not be detected in 32.7, 20, and 14.6% using the NucliSens, Versant, or TaqMan tests, respectively, suggesting potential viral load underestimation of some samples by all techniques. Only 32 (58.2%) samples from naive subjects were quantified by the three methods; the NucliSens test provided the highest HIV RNA values (mean, 4.87 log copies/ml), and the Versant test provided the lowest (mean, 4.16 log copies/ml). Viremia differences of greater than 1 log were seen in 8 (14.5%) of 55 specimens, occurring in 10.9, 7.3, and 5.4%, respectively, of the specimens in comparisons of Versant versus NucliSens, Versant versus TaqMan, and TaqMan versus NucliSens. Differences greater than 0.5 log, considered significant for clinicians, occurred in 45.5, 27.3, and 29% when the same assays were compared. Some HIV-1 strains, such as subtype G and CRF02_AG, showed more discrepancies in distinct quantification methods than others. In summary, an adequate design of primers and probes is needed for optimal quantitation of plasma HIV-RNA in non-B subtypes. Our data emphasize the need to use the same method for monitoring patients on therapy and also the convenience of HIV-1 subtyping.

Publisher

American Society for Microbiology

Subject

Microbiology (medical)

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3