Rapid bioluminescence method for bacteriuria screening

Author:

Pezzlo M T1,Ige V1,Woolard A P1,Peterson E M1,de la Maza L M1

Affiliation:

1. Department of Pathology, University of California Irvine Medical Center, Orange 92668.

Abstract

A study was performed to evaluate the UTIscreen (Los Alamos Diagnostics, Los Alamos, N. Mex.), a rapid bioluminescence bacteriuria screen. The UTIscreen was compared with three other rapid bacteriuria screens: the Bac-T-Screen (Vitek Systems, Hazelwood, Mo.), an automated filtration device; the Chemstrip LN (Boehringer Mannheim Diagnostics, BioDynamics, Indianapolis, Ind.), an enzyme dipstick; and the Gram stain. A semiquantitative plate culture was used as the reference method. Of the 1,000 specimens tested, 276 had colony counts of greater than 10(5) CFU/ml by the culture method. Of these, the UTIscreen detected 96% (265 of 276) using greater than or equal to 5% of the integrated light output of the standard reading as a positive interpretive breakpoint, the Bac-T-Screen detected 96% (266 of 276), the Chemstrip LN detected 90% (249 of 276), and the Gram stain detected 96% (264 of 276). Of the 214 probable pathogens isolated at greater than 10(5) CFU/ml, the UTIscreen detected 95% (204 of 214), the Bac-T-Screen detected 98% (210 of 214), the Chemstrip LN detected 92% (198 of 214), and the Gram stain detected 98% (209 of 214). The predictive values of negative test results at greater than 10(5) CFU/ml for the UTIscreen, the Bac-T-Screen, the Chemstrip LN, and the Gram stain were 98, 97, 93, and 98%, respectively. The overall specificities at greater than 10(5) CFU/ml for the UTIscreen, the Bac-T-Screen, the Chemstrip LN, and the Gram stain were 70, 48, 51, and 69%, respectively. There were 532 specimens with colony counts of >10(3) CFU/ml, and of these, the UTIscreen, the Bac-T-Screen, the Chemstrip LN, and the Gram stain detected 72, 81, 76, and 73%, respectively. Of the 249 probable pathogens isolated at >10(3) CFU/ml, the UTIscreen, the Bac-T-Screen, the Chemstrip LN, and the Gram stain detected 91, 95, 89, and 93%, respectively. The overall specificities at > 10(3) CFU/ml for these methods were 79, 55, 57, and 78%, respectively. The cost per test for detection was approximately $0.50 for the Chemstrip LN. Overall, the UTIscreen is rapid and easy to perform; its sensitivity compared favorably with those of the other screening methods; it had higher specificity than the Bac-T-Screen and Chemstrip LN; and it allowed for bathing of specimen.

Publisher

American Society for Microbiology

Subject

Microbiology (medical)

Reference25 articles.

1. Accuracy of calibratedloop transfer;Albers A. C.;J. Clin. Microbiol.,1983

2. Clinical evaluation of three rapid methods for the detection of significant bacteriuria;Bixier-Forell E.;J. Clin. Microbiol.,1985

3. Clarridge J. E. M. T. Pezzlo and K. L. Vosti. 1987. Cumitech 2A Laboratory diagnosis of urinary tract infections. Coordinating ed. A. L. Weissfeld. American Society for Microbiology Washington D.C.

4. College of American Pathologists Workload Recording Committee. 1987. Manual for laboratory workload recording method 1988 ed. p. 111-125. College of American Pathologists Traverse City Mich.

5. Comparison of the Lumac and Monolight systems for detection of bacteriuria by bioluminescence;Drow D. L.;J. Clin. Microbiol.,1984