Parallel comparison of accuracy of API 20E, Vitek GNI, MicroScan Walk/Away Rapid ID, and Becton Dickinson Cobas Micro ID-E/NF for identification of members of the family Enterobacteriaceae and common gram-negative, non-glucose-fermenting bacilli

Author:

O'Hara C M1,Tenover F C1,Miller J M1

Affiliation:

1. Nosocomial Pathogens Laboratory Branch, Centers for Disease Control and Prevention (C16), Atlanta, Georgia 30333.

Abstract

We compared the API 20E (21 h) (API; bioMérieux Vitek, Hazelwood, Mo.), the Vitek GNI card (4 to 18 h) (Vitek; bioMérieux Vitek), the identification portion of the MicroScan Walk/Away Rapid Neg Combo 3 panel (2 h) (W/A; Baxter Diagnostics, Inc., West Sacramento, Calif.), and the Becton Dickinson Cobas Micro ID-E/NF rotor (21 h) (Cobas; Becton Dickinson Diagnostic Instrument Systems, Sparks, Md.), versus conventional biochemicals for their abilities to identify accurately 252 strains of biochemically typical and atypical members of the family Enterobacteriaceae and common non-glucose-fermenting gram-negative bacilli. All strains used were included in the data base of each product. At the end of the initial incubation, 194 (77.0%), 213 (84.5%), 198 (78.6%), and 192 (76.2%) strains were correct to the genus and species levels with the API, Vitek, W/A, and Cobas systems, respectively. After additional biochemical tests were performed, as directed by each manufacturer's protocol, the numbers of strains correctly identified to the genus and species levels were 241 (95.6%), 234 (92.8%), 243 (96.4%), and 230 (91.3%) with the four systems, respectively. The errors were random in all systems, with the exception of two atypical Salmonella enteritidis strains, each of which was misidentified by three systems. After the initial recommended incubation period, both API and Cobas were significantly less accurate than Vitek (Yates' corrected P < 0.05). No significant differences were noted between the results of Vitek and W/A or between the results of API and W/A. After additional tests were completed, Cobas was significantly less accurate than W/A (P < 0.05) but was equal in accuracy to Vitek and API. API, Vitek, and W/A were equal in accuracy after these same additional tests. All four systems were significantly more accurate after additional biochemical testing than after the initial reporting period (194 of 252 versus 241 of 252 for API, 213 of 252 versus 234 of 252 for Vitek, 198 of 252 versus 243 or 252 for W/A, and 192 of 252 versus 230 of 252 for Cobas [P<0.05]).

Publisher

American Society for Microbiology

Subject

Microbiology (medical)

Reference16 articles.

1. Colonna P. 1990. Abstr. Annu. Meet. Am. Soc. Microbiol. 1990 C157 p. 370.

2. Biochemical identification of new species and biogroups of Enterobacteriaceae isolated from clinical specimens;Farmer J. J.;J. Clin. Microbiol.,1985

3. Evaluation of a new commercial system for the identification of Enterobacteriaceae and non-fermentative bacteria;Geiss H. K.;Eur. J. Clin. Microbiol. Infect. Dis.,1992

4. Evaluation of the autoSCAN-W/A system for rapid (2-hour) identification of members of the family Enterobacteriaceae;O'Hara C. M.;J. Clin. Microbiol.,1992

5. Comparison of the Cobas Micro with ID-E/NF rotor to conventional biochemicals for identification of Enterobacteriaceae and common glucose non-fermenters;O'Hara C. M.;Med. Microbiol. Lett.,1992