Abstract
ABSTRACT
The BACTEC MGIT 960 instrument is a fully automated system that exploits the fluorescence of an oxygen sensor to detect growth of mycobacteria in culture. Its performance was compared to those of the radiometric BACTEC 460 instrument and egg-based Lowenstein-Jensen medium. An identical volume of sample was inoculated in different media, and incubation was carried out for 6 weeks with the automatic systems and for 8 weeks on solid media. A total of 2,567 specimens obtained from 1,631 patients were cultured in parallel. Mycobacteria belonging to nine different taxa were isolated by at least one of the culture systems, with 75% of them being represented by
Mycobacterium tuberculosis
complex. The best yield was obtained with the BACTEC 460 system, with 201 isolates, in comparison with 190 isolates with the BACTEC MGIT 960 system and 168 isolates with Lowenstein-Jensen medium. A similar but not significant difference was obtained when the most-represented organisms, the
M. tuberculosis
complex,
Mycobacterium xenopi
, and the
Mycobacterium avium
complex, were analyzed separately and when combinations of a solid medium with the BACTEC MGIT 960 system and with the BACTEC 460 system were considered. The shortest times to detection were obtained with the BACTEC MGIT 960 system (13.3 days); 1.5 days earlier than that with the BACTEC 460 system (14.8 days) and 12 days earlier than that with Lowenstein-Jensen medium (25.6 days). The BACTEC MGIT 960 system had a contamination rate of 10.0%, intermediate between those of the radiometric system (3.7%) and the egg-based medium (17.0%). We conclude, therefore, that the BACTEC MGIT 960 system is a fully automated, nonradiometric instrument that is suitable for the detection of growth of tuberculous and other mycobacterial species and that is characterized by detection times that are even shorter than that of the “gold standard,” the BACTEC 460 system. The contamination rate was higher than that for the radiometric BACTEC 460 system and needs to be improved.
Publisher
American Society for Microbiology
Reference18 articles.
1. Essential components of a tuberculosis prevention and control program;Centers for Disease Control and Prevention;Morbid. Mortal. Weekly Rep.,1995
2. Multicenter Evaluation of the BACTEC MGIT 960 System for Recovery of Mycobacteria
3. Mycobacteriology laboratory;Heifets L.;Clin. Chest Med.,1997
4. Nolte
F. S.
Metchock
B. Mycobacterium Manual of clinical microbiology
6th ed.
Murray
P. R.
Baron
E. J.
Pfaller
M. A.
Tenover
F. C.
Yolken
R. H.
1995
400
437
ASM Press
Washington D.C
5. Application of new technology to the detection, identification, and antimicrobial susceptibility testing of mycobacteria;Pfaller M. A.;Am. J. Clin. Pathol.,1994
Cited by
184 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献