Collins M T,Weaver N,Ellis R P
Abstract
Fifty serotyped isolates each of Pasteurella multocida and Pasteurella haemolytica were tested on the API 20E strip (Analytab Products, Plainview, N.Y.), the Oxi/Ferm tube (Roche Diagnostics, Nutley, N.J.), and the Minitek system (BBL Microbiology Systems, Cockeysville, Md.). None of the rapid test systems reliable identified these organisms. With the API system, discrepancies between expected and actual results for the oxidase test and nitrate test frequently resulted in misidentification or no identification. The Minitek system misidentified 68% of the P. haemolytica isolates. The Minitek identification of Pasteurella depends on 100% positive xylose reactions, whereas only 56% of the P. haemolytica strains were positive for xylose fermentation. The Oxy/Ferm system, instead of giving a definitive identification, in most instances merely placed Pasteurella in a category of similar organisms.
Publisher
American Society for Microbiology
Reference19 articles.
1. Four methods for identification of gram-negative nonfermenting rods: organisms more commonly encountered in clinical specimens;Applebaum P. C.;J. Clin. Microbiol.,1980
2. A clinical evaluation of the API microtube system for identification of;Brooks K. S.;Enterobacteriaecae. Am. J. Med. Technol.,1974
3. Evaluation of two rapid methods for identification of commonly encountered nonfermenting or oxidase-positive, gram-negative rods;Dowda H.;J. Clin. Microbiol.,1977
4. Clinical evaluation of the Minitek differential system for identification of Enterobacteriaceae;Finklea P. J.;J. Clin. Microbiol.,1977
5. Rapid plate agglutination procedure for serotyping Pasteurella haemolytica;Frank G. H.;J. Clin. Microbiol.,1978
Cited by
17 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献