Affiliation:
1. Leeds Biofilm Group, Department of Microbiology, Leeds Teaching Hospitals, Leeds LS29NL, United Kingdom
Abstract
ABSTRACT
This biofilm study was conducted to assess the in vitro activity of tetrasodium EDTA on catheters that had been routinely removed from hemodialysis patients at Leeds Teaching Hospitals Trust due to maturation of fistula. Catheters were screened by culture of through-catheter flush, and isolates were identified by standard methodologies; 20 isolates were found to be biofilm positive. Initial biofilm cell count levels averaged above 10
5
CFU/1-cm catheter section. Bacteria identified in the biofilms were gram-negative (1 isolate), gram-positive (11 isolates), or mixed species (8 isolates). After a 24-h lock, 40 mg of tetrasodium EDTA per ml was effective at eradicating the total biofilm viable count in almost all cases. The efficacy of tetrasodium EDTA as a catheter lock potentially shows that this agent could substantially reduce catheter-related infections and be used to treat patients with limited access.
Publisher
American Society for Microbiology
Reference15 articles.
1. Anaissie, E., G. Samonis, D. Kontoyiannis, J. Costerton, U. Sabharwal, G. Bodey, and I. Raad. 1995. Role of catheter colonization and infrequent hematogenous seeding in catheter-related infections. Eur. J. Clin. Microbiol. Infect. Dis.14:135-137.
2. Randomized, Double-Blind Trial of an Antibiotic-Lock Technique for Prevention of Gram-Positive Central Venous Catheter-Related Infection in Neutropenic Patients with Cancer
3. Christensen G. D. L. Baldassari and W. A. Simpson. 1994. Colonization of medical devices by coagulase-negative staphylococci p. 45-78. In A. L. Bisno and F. A. Waldvogel (ed.) Infections associated with indwelling medical devices 2nd ed. American Society for Microbiology Washington D.C.
4. Effect of vancomycin hydrochloride on Staphylococcus epidermidis biofilm associated with silicone elastomer
5. Garland, J. S., K. Henrickson, and D. G. Maki. 2002. The 2002 Hospital Infection Control Practices Advisory Committee, Centers for Disease Control and Prevention, guideline for prevention of intravascular device-related infection. Pediatrics110:1009-1013.