Evaluation of Six Commercial Point-of-Care Tests for Diagnosis of Acute Dengue Infections: the Need for Combining NS1 Antigen and IgM/IgG Antibody Detection To Achieve Acceptable Levels of Accuracy

Author:

Blacksell Stuart D.12,Jarman Richard G.3,Bailey Mark S.4,Tanganuchitcharnchai Ampai1,Jenjaroen Kemajittra1,Gibbons Robert V.3,Paris Daniel H.12,Premaratna Ranjan5,de Silva H. Janaka5,Lalloo David G.6,Day Nicholas P. J.12

Affiliation:

1. Mahidol-Oxford Tropical Medicine Research Unit (MORU), Faculty of Tropical Medicine, Mahidol University, 420/6 Rajvithi Road, Bangkok 10400, Thailand

2. Centre for Tropical Medicine, University of Oxford, Churchill Hospital, Oxford OX3 7LJ, England, United Kingdom

3. Armed Forces Research Institute of Medical Sciences, 315/6 Rajvithi Road, Bangkok 10400, Thailand

4. Department of Military Medicine, Royal Centre for Defence Medicine, Vincent Drive, Birmingham B15 2SQ, England, United Kingdom

5. Department of Medicine, Faculty of Medicine, University of Kelaniya, P.O. Box 6, Thalagolla Road, Ragama, Sri Lanka

6. Clinical Research Group, Liverpool School of Tropical Medicine, Pembroke Place, Liverpool L3 5QA, England, United Kingdom

Abstract

ABSTRACT Six assays were evaluated in this study to determine their suitability for the diagnosis of acute dengue infection using samples from 259 Sri Lankan patients with acute fevers (99 confirmed dengue cases and 160 patients with other confirmed acute febrile illnesses): (i) the Merlin dengue fever IgG & IgM combo device (Merlin), (ii) the Standard Diagnostics Dengue Duo nonstructural 1 (NS1) antigen and IgG/IgM combo device (Standard Diagnostics, South Korea), (iii) the Biosynex Immunoquick dengue fever IgG and IgM (Biosynex, France) assay, (iv) the Bio-Rad NS1 antigen strip (Bio-Rad, France), (v) the Panbio Dengue Duo IgG/IgM Cassette (Inverness, Australia), and (vi) the Panbio dengue NS1 antigen strip (Inverness, Australia). The median number of days of fever prior to admission sample collection was 5 days (interquartile range, 3 to 7 days). Sensitivity and specificity of the NS1 antigen tests ranged from 49 to 59% and from 93 to 99%, respectively, and sensitivity and sensitivity of the IgM antibody test ranged from 71 to 80% and from 46 to 90%, respectively. Combining the NS1 antigen and IgM antibody results from the Standard Diagnostics Dengue Duo test gave the best compromise of sensitivity and specificity (93% and 89%, respectively) and provided the best sensitivity in patients presenting at different times after fever onset. The Merlin IgM/IgG antibody tests correctly classified 64% and 86% of the primary and secondary dengue infection cases, respectively, and the Standard Diagnostics IgM/IgG antibody tests correctly classified 71% and 83% of the primary and secondary dengue infection cases, respectively. This study provides strong evidence of the value of combining dengue antigen- and antibody-based test results in the rapid diagnostic test (RDT) format for the acute diagnosis of dengue.

Publisher

American Society for Microbiology

Subject

Microbiology (medical),Clinical Biochemistry,Immunology,Immunology and Allergy