Abstract
ABSTRACT
The Agilent 2100 bioanalyzer (Agilent Technologies, Palo Alto, Calif.) utilizes capillary electrophoresis on a microchip device (LabChip 7500; Caliper Technologies, Mountain View, Calif.) that is capable of rapidly sizing small DNA fragments. To determine whether the system could replace conventional restriction fragment length polymorphism (RFLP) typing by agarose gel electrophoresis, we compared the analyzer with conventional flagellin RFLP for typing
Campylobacter jejuni
. Ninety-seven isolates representing 46 Fla types were initially analyzed. Correct Fla types were detected in 59% of the isolates. The major problem with the system was in resolving samples containing multiple DNA fragments differing from 8 to 20 bp. Overall, the bioanalyzer has the potential to replace conventional RFLP analysis by gel electrophoresis, but improvements in the chip separation are needed.
Publisher
American Society for Microbiology
Reference8 articles.
1. Arbeit
R. D.
Laboratory procedures for the epidemiologic analysis of microorganisms
Manual of clinical microbiology
7th ed.
Murray
P. R.
Baron
E. J.
Pfaller
M. A.
Tenover
F. C.
Yolken
R. H.
1999
116
137
ASM Press
Washington D.C.
2. Safer and healthier foods;Centers for Disease Control and Prevention;Morb. Mortal. Wkly. Rep.,1999
3. Flagellin gene typing of Campylobacter jejuni by restriction fragment length polymorphism analysis
4. Analysis of HL and O serotypes of Campylobacter strains by the flagellin gene typing system
5. Newell
D. G.
Frost
J. A.
Duim
B.
Wagenaar
J. A.
Madden
R. H.
van der Plas
J.
On
S. L. W.
New developments in the subtyping of Campylobacter species
Campylobacter.
Nachamkin
I.
Blaser
M. J.
2000
27
44
ASM Press
Washington D.C.
Cited by
63 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献