Comparison of Three Different FDA-Approved Plasma HIV-1 RNA Assay Platforms Confirms the Virologic Failure Endpoint of 200 Copies per Milliliter Despite Improved Assay Sensitivity

Author:

Lalama Christina M.1,Jennings Cheryl2,Johnson Victoria A.3,Coombs Robert W.4,McKinnon John E.5,Bremer James W.2,Cobb Bryan R.6,Cloherty Gavin A.7,Mellors John W.8,Ribaudo Heather J.1

Affiliation:

1. Harvard T.H. Chan School of Public Health, Boston, Massachusetts, USA

2. Rush University Medical Center, Chicago, Illinois, USA

3. Birmingham Veterans Affairs Medical Center and University of Alabama at Birmingham School of Medicine, Birmingham, Alabama, USA

4. University of Washington, Seattle, Washington, USA

5. Henry Ford Hospital, Detroit, Michigan, USA

6. Roche Molecular Systems, Pleasanton, California, USA

7. Abbott Molecular, Des Plaines, Illinois, USA

8. University of Pittsburgh, Pittsburgh, Pennsylvania, USA

Abstract

ABSTRACT Discrepancies between HIV-1 RNA results assayed by different FDA-approved platforms have been reported. Plasma samples collected from 332 randomly selected clinical trial participants during the second year of antiretroviral treatment were assayed with three FDA-approved platforms: UltraSensitive Roche Amplicor Monitor, v1.5 (Monitor), the Abbott RealTime HIV-1 test on the m2000 system (Abbott), and the Roche TaqMan HIV-1 test, v2.0 (TaqMan). Samples from 61 additional participants with confirmed HIV-1 RNA levels of >50 copies/ml during trial follow-up were also included. Endpoints were HIV-1 RNA quantification of ≤50 copies/ml versus >50 copies/ml at an individual-sample level (primary) and determination of confirmed virologic failure (VF) from longitudinal samples. A total of 389 participants had results obtained from all assays on at least one sample (median = 6). Proportions of results of >50 copies/ml were 19% (Monitor), 22% (TaqMan), and 25% (Abbott). Despite indication of strong agreement (Cohen's kappa, 0.76 to 0.82), Abbott was more likely to detect HIV-1 RNA levels of >50 copies/ml than Monitor (matched-pair odds ratio [mOR] = 4.2; modified Obuchowski P < 0.001) and TaqMan (mOR = 2.1; P < 0.001); TaqMan was more likely than Monitor (mOR = 2.6; P < 0.001). Despite strong agreement in classifying VF across assay comparisons (kappa, 0.75 to 0.92), at a 50-copies/ml threshold, differences in the probability of VF classification (in the same direction as primary) were apparent (all McNemar's P < 0.007). At a 200-copies/ml VF threshold, no differences between assays were apparent (all P > 0.13). Despite strong agreement among assays, significant differences were observed with respect to detecting HIV-1 RNA levels of >50 copies/ml and identifying VF at the 50-copies/ml threshold. This has important implications for the definition of VF in clinical trials and clinical practice.

Publisher

American Society for Microbiology

Subject

Microbiology (medical)

Reference17 articles.

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3