Operational Feasibility of Using Loop-Mediated Isothermal Amplification for Diagnosis of Pulmonary Tuberculosis in Microscopy Centers of Developing Countries

Author:

Boehme Catharina C.1,Nabeta Pamela2,Henostroza German2,Raqib Rubhana3,Rahim Zeaur3,Gerhardt Martina4,Sanga Erica4,Hoelscher Michael5,Notomi Tsugunori6,Hase Tetsu6,Perkins Mark D.1

Affiliation:

1. Foundation for Innovative New Diagnostics, Geneva, Switzerland

2. Universidad Peruana Cayetano Heredia, Lima, Peru

3. International Centre for Diarrheal Diseases Research, Dhaka, Bangladesh

4. Mbeya Medical Research Programme, Mbeya, Tanzania

5. Department of Infectious Diseases and Tropical Medicine, University of Munich, Munich, Germany

6. Eiken Chemical Co. Ltd., Tokyo, Japan

Abstract

ABSTRACT The characteristics of loop-mediated isothermal amplification (LAMP) make it a promising platform for the molecular detection of tuberculosis (TB) in developing countries. Here, we report on the first clinical evaluation of LAMP for the detection of pulmonary TB in microscopy centers in Peru, Bangladesh, and Tanzania to determine its operational applicability in such settings. A prototype LAMP assay with simplified manual DNA extraction was evaluated for accuracy and ease of use. The sensitivity of LAMP in smear- and culture-positive sputum specimens was 97.7% (173/177 specimens; 95% confidence interval [CI], 95.5 to 99.9%), and the sensitivity in smear-negative, culture-positive specimens was 48.8% (21/43 specimens; CI, 33.9 to 63.7%). The specificity in culture-negative samples was 99% (500/505 specimens; CI, 98.1 to 99.9%). The average hands-on time for testing six samples and two controls was 54 min, similar to that of sputum smear microscopy. The optimal amplification time was 40 min. No indeterminate results were reported, and the interreader variability was 0.4%. Despite the use of a single room without biosafety cabinets for all procedures, no DNA contamination was observed. The assay was robust, with high end-point stability and low rates of test failure. Technicians with no prior molecular experience easily performed the assay after 1 week of training, and opportunities for further simplification of the assay were identified.

Publisher

American Society for Microbiology

Subject

Microbiology (medical)

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3