Author:
Barry A L,Coyle M B,Thornsberry C,Gerlach E H,Hawkinson R W
Abstract
Standard Bauer-Kirby disk tests were performed with 85 selected isolates, each tested in triplicate by four different investigators. Each disk test was observed, and zone diameters were measured, under two lighting conditions (transmitted light and reflected light). The two lighting systems produced similar zone measurements (+/-2 mm) with 96% of the tests. When there were greater differences, zones appeared to be larger when observed with reflected light. Interlaboratory reproducibility was much greater when using reflected light rather than transmitted light. We concluded that zone diameters should be measured from the back of the plate while it is resting on, or held 2 to 3 inches [ca. 5.1 to 7.6 cm] above, a black, nonreflecting, flat surface, illuminated by a reflected light source.
Publisher
American Society for Microbiology
Reference3 articles.
1. Antibiotic susceptibility testing by a standardized single disc method;Bauer A. W.;Am. J. Clin. Pathol.,1966
2. National Committee for Clinical Laboratory Standards. 1975. Performance standards for antimicrobial disc susceptilibity tests. National Committee for Clinical Laboratory Standards Villanova Pa.
3. Searle S. R. 1971. Linear models p. 376-420. John Wiley and Sons Inc. New York.
Cited by
101 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献