Author:
Hunter E F,Adams M R,Orrison L H,Pender B J,Larsen S A
Abstract
Immunofluorescent staining of Treponema pallidum was studied to clarify the effect of three factors on the results of the fluorescent treponemal antibody-absorption test: (i) heat inactivation of sera at 56 degrees C for 30 min before testing, (ii) use of multicircle slides, and (iii) tungsten illumination to visualize and assess unstained treponemes on reactive as well as nonreactive smears. It was found that serum inactivation before testing was not necessary for detection of immunoglobin G antibody, but an immunoglobulin M prozone was detected in unheated serum. On multicircle slides, it was demonstrated that a false-positive reaction could be obtained in 30 s at 37 and 25 degrees C if a smear where a nonreactive serum had been placed was crossed by a strongly reactive serum from another circle. Tungsten illumination proved necessary for correct assessment of unstained treponemes on all fluorescent treponemal antibody-aborption test smears, reactive or nonreactive. The possible role of these factors in incorrect fluorescent treponemal antibody-absorption test results is discussed.
Publisher
American Society for Microbiology
Cited by
2 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献