Affiliation:
1. Department of Pathology, University of Iowa College of Medicine, Iowa City 52242.
Abstract
Preliminary interpretive zone diameter criteria were calculated for the 5-micrograms cefdinir disk diffusion test by using two potential MIC breakpoints (less than or equal to 0.5 and less than or equal to 1 micrograms/ml). The absolute agreement between tests ranged from 85.9 to 92.4%, and the false-susceptibility errors were principally contributed by the Enterobacter spp. (2.2% error). One proposed criterion was greater than or equal to 20-mm zone diameter (less than or equal to 1 micrograms/ml) for susceptibility and less than or equal to 16-mm zone diameter (greater than 2 micrograms/ml) for resistance to cefdinir. Clinical laboratory users of the disk diffusion method should be cautioned about the possibility of very major interpretive errors among enterobacter isolates.
Publisher
American Society for Microbiology
Cited by
12 articles.