McCracken A W,Martin W J,McCarthy L R,Schwab D A,Cooper B H,Helgeson N G,Prowant S,Robson J
Abstract
The precision, accuracy, and other performance characteristics of the MS-2 (Abbott Laboratories, Diagnostic Division, Dallas, Tex.) system for the identification of Enterobacteriaceae were evaluated in a collaborative study involving three clinical laboratories. When identifying 150 unknown, coded organisms, the MS-2 system was 97%, 98%, and 93% accurate, respectively, in three laboratories. The system showed an overall accuracy of 94% when compared with conventional manual tube methods in identifying 1,154 clinical isolates of 26 species of Enterobacteriaceae. Discrepancies between automated and conventional methods were chiefly caused by biochemical variants, especially among Enterobacter species. The MS-2 system was rapid and simple to operate and produced printed results of bacterial identification in 5 h. The cost of disposable components compared favorably with commercial, visually read systems for identifying Enterobacteriaceae.
Publisher
American Society for Microbiology