Affiliation:
1. Department of Pathology and Area Laboratory Services, Brooke Army Medical Center, Fort Sam Houston, Texas 78234
Abstract
The Oxi/Ferm test system was evaluated for accuracy and reliability for identification of nonfermentative and oxidase-positive fermentative bacteria by using 375 bacterial strains obtained from stock culture and clinical specimens. The Oxi/Ferm system is a compartmentalized tube containing eight media to provide nine biochemical test results. When combined with the oxidase test, the results corresponding to the positive reactions are totaled and the composite number is located in the coding manual to identify the organisms. The 375 isolates studied were evaluated for accuracy of identification, using both the original and revised code manuals. In comparison with the conventional media used, there was 100% correlation in tests for hydrogen sulfide and indole production, over 96% for nitrogen gas, arginine, and urease, over 92% for xylose and dextrose oxidation, and less than 90% for citrate utilization and dextrose fermentation. There was an overall accuracy in identification of 89.3% using the original manual, with accuracy revised slightly upward to 90.7% using the revised manual. There was 100% accuracy in identification with 44.0% of the strains tested (11 species) using the original manual and with 66.1% (16 species) using the revised manual. Thirteen of the 40 original misidentifications and 14 of 35 revised misidentifications resulted from failure to code and were unidentifiable by Oxi/Ferm. The remainder were incorrectly identified or could not be differentiated from closely related strains. Eleven strains of
Alcaligenes odorans
were correctly identified using the original code, whereas no code was provided in the revised manual. The Oxi/Ferm system is both simple and rapid and is satisfactory for identification of the more common isolates.
Publisher
American Society for Microbiology
Reference16 articles.
1. Antibiotic su$ceptibility testing by a standardized single disc method;Bauer A. W.;Am. J. Clin. Pathol.,1966
2. Comparison of the R-B system and the Enterotube for the identification of;Coppel S. P.;Enterobacteriaceae. Am. J. Clin. Pathol.,1974
3. Motility-indole-lysine-sulfide medium;Ederer G. M.;J. Chin. Microbiol.,1975
4. Edwards P. R. and W. H. Ewing. 1972. Identification of Enterobacteriaceae. 3rd ed. Burgess Publishing Co. Minneapolis.
5. Multibiochemical test system for distinguishing enteric and other gram-negative bacilli;Elston J. R.;Appl. Microbiol.,1971
Cited by
16 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献