Abstract
ABSTRACT
Recently, the Centers for Disease Control and Prevention reported an accurate, sensitive, specific, reproducible, and quantitative enzyme-linked immunosorbent assay (ELISA) for immunoglobulin G (IgG) antibodies to
Bacillus anthracis
protective antigen (PA) in human serum (C. P. Quinn, V. A. Semenova, C. M. Elie et al., Emerg. Infect. Dis. 8:1103-1110, 2002). The ELISA had a minimum detectable concentration (MDC) of 0.06 μg/ml, which, when dilution adjusted, yielded a whole-serum MDC of 3.0 μg of anti-PA IgG per ml. The reliable detection limit (RDL) was 0.09 μg/ml, while the dynamic range was 0.06 to 1.7 μg/ml. The diagnostic sensitivity of the assay was 97.6% and the diagnostic specificity was 94.2% for clinically verified cases of anthrax. A competitive inhibition anti-PA IgG ELISA was also developed to enhance the diagnostic specificity to 100%. We report a newly developed fluorescence covalent microbead immunosorbent assay (FCMIA) for
B. anthracis
PA which was Luminex xMap technology. The FCMIA MDC was 0.006 μg of anti-PA IgG per ml, the RDL was 0.016 μg/ml, and the whole-serum equivalent MDC was 1.5 μg/ml. The dynamic range was 0.006 to 6.8 μg/ml. Using this system, we analyzed 20 serum samples for anti-PA IgG and compared our results to those measured by ELISA in a double-masked analysis. The two methods had a high positive correlation (
r
2
= 0.852;
P
< 0.001). The FCMIA appears to have benefits over the ELISA for the measurement of anti-PA IgG, including greater sensitivity and speed, enhanced dynamic range and reagent stability, the use of smaller sample volumes, and the ability to be multiplexed (measurement of more than one analyte simultaneously), as evidenced by the multiplexed measurement in the present report of anti-PA and anti-lethal factor IgG in serum from a confirmed clinical anthrax infection.
Publisher
American Society for Microbiology
Subject
Microbiology (medical),Clinical Biochemistry,Immunology,Immunology and Allergy
Reference19 articles.
1. Bardsley, W. G., and P. B. McGinlay. 1987. The use of non-linear regression analysis and the F test for model discrimination with dose-response curves and ligand binding data. J. Theor. Biol.126:183-201.
2. Biagini, R. E., D. M. Murphy, D. L. Sammons, J. P. Smith, C. A. Striley, and B. A. MacKenzie. 2002. Development of multiplexed fluorescence microbead covalent assays (FMCAs) for pesticide biomonitoring. Bull. Environ. Contam. Toxicol.68:470-477.
3. Biagini, R. E., S. A. Schlottmann, D. L. Sammons, J. P. Smith, J. C. Snawder, C. A. Striley, B. A. MacKenzie, and D. N. Weissman. 2003. Method for simultaneous measurement of antibodies to 23 pneumococcal capsular polysaccharides. Clin. Diagn. Lab. Immunol.10:744-750.
4. Simultaneous Detection of 15 Human Cytokines in a Single Sample of Stimulated Peripheral Blood Mononuclear Cells
5. Hornung, R., and L. Reed. 1990. Estimate of average concentration in the presence of non-detectable values. Appl. Occup. Environ. Hyg.5:46-51.