Abstract
ABSTRACT
A simplified lateral-flow assay for the detection of antibodies to HIV using magnetic-bead conjugates and multibranched peptides from both HIV-1 and HIV-2 was developed. Magnetic immunochromatography testing (MICT) uses a standard lateral-flow platform that incorporates magnetic-bead conjugates for quantitative measurement of the magnetic field distortion associated with the bound magnetic conjugate (reported as adjusted relative magnetic units [MAR]). The results of the optimized MICT assay were compared to standard enzyme immunoassay (EIA) and Western blotting (WB) results using a blinded 649-member panel of specimens from the United States, Cameroon, and West Africa. The panel was comprised of samples from individuals infected with various HIV-1 subtypes (
n
= 234) or HIV-2 (
n
= 65) and HIV-seronegative specimens (
n
= 350). Additionally, 13 HIV-1 seroconversion panels (total specimens = 85), a worldwide panel containing seven of the major circulating HIV-1 subtypes (
n
= 18), an HIV-2 panel, an HIV-1/HIV-2 mixed panel, and 100 prospective specimens were tested with completely concordant results. Assay reproducibility (observed MAR) for both intra- and interrun testing was excellent, with coefficients of variation of <12%. MICT can provide a rapid, low-cost method of determining HIV antibody status requiring no subjective interpretations.
Publisher
American Society for Microbiology
Subject
Microbiology (medical),Clinical Biochemistry,Immunology,Immunology and Allergy
Reference27 articles.
1. HIV simple/rapid assays: operational characteristics 12. 2002
2. Brattegaard, K., J. Kouadio, M. L. Adom, R. Doorly, J. R. George, and K. M. De Cock. 1993. Rapid and simple screening and supplemental testing for HIV-1 and HIV-2 infections in west Africa. AIDS7:883-885.
3. Centers for Disease Control and Prevention. 2008. False-positive oral fluid rapid HIV tests—New York City, 2005-2008. MMWR Morbid. Mortal. Wkly. Rep.57:660-665.
4. Constantine, N. T., L. Zekeng, A. K. Sangare, L. Gurtler, R. Saville, H. Anhary, and C. Wild. 1997. Diagnostic challenges for rapid human immunodeficiency virus assays. Performance using HIV-1 group O, HIV-1 group M, and HIV-2 samples. J. Hum. Virol.1:45-51.
5. De Cock, K., A. Porler, J. Koudadio, et al. 1991. Cross-reactivity on Western blots in HIV-1 and HIV-2 infections. AIDS5:859-863.
Cited by
27 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献