Evaluation of the Performance of the Automated NucliSENS easyMAG and EasyQ Systems versus the Roche AmpliPrep-AMPLICOR Combination for High-Throughput Monitoring of Human Immunodeficiency Virus Load

Author:

Stevens W.1,Horsfield P.1,Scott L. E.1

Affiliation:

1. Department of Molecular Medicine and Hematology, University of the Witwatersrand and the National Health Laboratory Service, Johannesburg, South Africa

Abstract

ABSTRACT This study presents the data of an evaluation of the automated Nuclisens easyMAG and EasyQ systems versus the Roche AmpliPrep-AMPLICOR combination for testing of high-volume human immunodeficiency virus (HIV) load. This represents a follow-up of a previous study investigating the performance of the real-time Nuclisens assay using the semiautomated NucliSENS miniMAG extraction procedure. Three hundred eighteen patient samples were analyzed using both methods. The easyMAG-EasyQ HIV type 1 system has a higher sensitivity and broader dynamic range than the Cobas AmpliPrep-AMPLICOR system when the standard Roche assay is used alone, 25 to 3,000,000 IU/ml versus 400 to 750,000 HIV RNA copies/ml, respectively. There was significant correlation between the assays (0.93; P < 0.0001), with good accuracy (percent similarity mean μ = 96%), good precision (percent similarity standard deviation = 4.97%), and overall good agreement with a low percent similarity coefficient of variation of 5.17 to 6.11%. Bland-Altman analysis revealed that the AMPLICOR assay generated higher values than the EasyQ combination, with 95% of results within clinically acceptable limits. The throughput of samples was greatly improved using the easyMAG-EasyQ system, allowing 144 samples to be processed within 6 h. The potential for contamination has been dramatically reduced using the automated extraction system. Additional negative controls have been added to the kit to monitor for contamination based on the South African experience. This assay thus presents a real option for monitoring HIV load assays in high-volume testing environments.

Publisher

American Society for Microbiology

Subject

Microbiology (medical)

Reference7 articles.

1. Bland, J. M., and D. G. Altman. 1986. Statistical methods for assessing agreement between two methods of clinical measurement. Lanceti:307-310.

2. de Mendoza, C., M. Koppelman, B. Montes, V. Ferre, V. Soriano, H. Cuypers, M. Segondy, and T. Oosterlaken. 2005. Multicenter evaluation of the NucliSens EasyQ HIV-1 v1.1 assay for the quantitative detection of HIV-1 RNA in plasma. J. Virol. Methods127:54-59.

3. National Department of Health. 2004. National antiretroviral treatment guidelines. National Department of Health Johannesburg South Africa. http://www.doh.gov.za/docs/factsheets/guidelines/artguide04-f.html .

4. Scott, L. E., J. S. Galpin, and D. K. Glencross. 2003. Multiple method comparison: statistical model using percentage similarity. Cytometry B Clin. Cytom.54:46-53.

5. Stevens, W., T. Wiggill, P. Horsfield, L. Coetzee, and L. E. Scott. 2005. Evaluation of the NucliSens EasyQ assay in HIV-1-infected individuals in South Africa. J. Virol. Methods124:105-110.