Abstract
Two rapid urine screening tests, the Chemstrip LN (BioDynamics, Indianapolis, Ind.) and the Bac-T-Screen urine screening device (Marion Laboratories, Inc., Kansas City, Mo.), were evaluated as techniques to predict bacteriuria as quantitated by either conventional culture or the AutoMicrobic system (Vitek Systems, Inc., Hazelwood, Mo.). A total of 666 urine specimens were analyzed by both screening tests as well as the AutoMicrobic system and quantitative culture. The sensitivities of both Chemstrip LN and Bac-T-Screen for the detection of low levels of bacteriuria (greater than or equal to 10(3) CFU/ml) were comparable (73.3 and 74.4%, respectively) and were too low to recommend their use as a primary urine screen. Their excellent predictive value of a negative result at the 10(5) CFU/ml level (96 and 97.5%, respectively) makes them potentially useful in predicting urine specimens with less than 10(5) CFU/ml. The use of either of these tests in combination with the AutoMicrobic system markedly decreased the time required to classify urine specimens. Their low cost relative to the AutoMicrobic system urine card makes the use of either test cost effective as a screen for the AutoMicrobic system.
Publisher
American Society for Microbiology
Reference23 articles.
1. Urine screening-to screen or not to screen;Bartlett R. C.;Clin. Microbiol. Newsl.,1984
2. Detection of bacteriuria by leukocyte esterase, nitrite, and the AutoMicrobic;Bartlett R. C.;System. Am. J. Clin. Pathol.,1984
3. Asymptomatic bacteriuria: the need for screening;Bates H. M.;Lab. Manage.,1982
4. Measurement of urinary leukocyte esterase activity: a screening test for urinary tract infections;Chernow B.;Ann. Emerg. Med.,1984
5. Colton T. 1974. Inference on proportions p. 174-177. Statistics in medicine. Little Brown & Co. Boston.
Cited by
20 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献