Multicenter clinical evaluation of a continuous monitoring blood culture system using fluorescent-sensor technology (BACTEC 9240)

Author:

Nolte F S1,Williams J M1,Jerris R C1,Morello J A1,Leitch C D1,Matushek S1,Schwabe L D1,Dorigan F1,Kocka F E1

Affiliation:

1. Clinical Microbiology Laboratory, Emory University Hospital, Atlanta, Georgia.

Abstract

The BACTEC 9240 (Becton Dickinson Diagnostic Instrument Systems, Sparks, Md.) is a new continuous-monitoring blood culture system that uses internal, fluorescent-CO2 sensors. In a multicenter clinical trial, organism yield and times to detection with the prototype BACTEC 9240 system were compared with those of the BACTEC NR 660 system. Equal volumes of blood were inoculated into the bottles included in the study blood culture sets (aerobic and anaerobic 9240 and NR6A and NR7A bottles). A total of 9,391 aerobic and 8,951 anaerobic bottle pairs were inoculated with 9,801 blood specimens. A total of 587 clinically significant positive blood cultures and 415 cases of sepsis were studied. The standard 9240 aerobic bottle detected significantly more Staphylococcus aureus (P < 0.05), coagulase-negative staphylococci (P < 0.01), and total microorganisms (P < 0.001) than the NR6A bottle. The standard 9240 anaerobic bottle detected significantly more coagulase-negative staphylococci (P < 0.001), members of the family Enterobacteriaceae (P < 0.01), and total microorganisms (P < 0.001) than the NR7A bottle. A total of 420 positive cultures were detected in both systems; for 284, the time to detection was equivalent with both systems (within 12 h); for 123, the 9240 system was faster; and for 13, the NR 660 system was faster (P < 0.001). The average times to detection for the 9240 and the NR 660 systems were 20.2 and 27.5 h, respectively. Ninety-nine cultures were positive only in the 9240 system, and 68 cultures were positive only in the NR 660 system (P < 0.02). The 9240 system also detected significantly more episodes of bacteremia (P < 0.001). The false-positive rates for the 9240 and NR 660 systems were 2.2 and 2.3%, respectively. The false-negative rates for the two systems after 5 days of incubation did not differ significantly. The contamination rates for the 9240 and NR 660 systems were 1.9 and 1.5%, respectively (P < 0.05). In conclusion, the prototype 9240 system detected more clinically significant positive blood cultures and did so sooner than the NR 660 system, with the additional advantages of full automation, continuous monitoring, and noninvasive sampling.

Publisher

American Society for Microbiology

Subject

Microbiology (medical)

Reference11 articles.

1. Comparison of recovery of organisms from blood cultures diluted 10% (volume/volume) and 20% (volume/volume);Aukenthaler R.;J. Clin. Microbiol.,1982

2. Becton Dickinson Diagnostic Instrument Systems. 1991. Package insert BACTEC NR6A NR7A and NR8A. Becton Dickinson Sparks Md.

3. Effect of sodium polyanethol sulfonate in blood cultures;Eng J.;J. Clin. Microbiol.,1975

4. Detection of positive blood cultures by the BACTEC NR660. The clinical importance of five versus seven days of testing;Masterson K. C.;Am. J. Clin. Pathol.,1988

5. McNemar Q. 1962. Psychological statistics 3rd ed. p. 209-239. John Wiley & Sons Inc. New York.

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3