Reader error in determining minimal inhibitory concentrations with microdilution susceptibility test panels

Author:

Barry A L,Braun L E

Abstract

Microdilution susceptibility test panels were independently read by two different technologists. One or more discrepancies between readers occurred with 5% of the 25,022 recorded minimal inhibitory concentrations, but only 0.4% of the tests varied by greater than or equal to 2 dilution intervals.

Publisher

American Society for Microbiology

Subject

Microbiology (medical)

Reference4 articles.

1. A system for reporting quantitative antimicrobial susceptibility test results;Barry A. L.;Am. J. Clin. Pathol.,1979

2. Hoeprich P. D. A. L. Barry and G. D. Fay. 1971. Synthetic medium for susceptibility testing p. 494-497. Antimicrob. Agents and Chemother. 1970.

3. On-line computer quality control of antibiotic-sensitivity testing;Petralli J.;N. Engl. J. Med.,1970

4. Susceptibility tests for sulfamethoxazole-trimethoprim by a broth microdilution procedure;Swenson J. M.;Curr. Microbiol.,1978

Cited by 13 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370