Author:
Jones C,MacPherson D W,Stevens D L
Abstract
The Chemstrip LN (Boehringer Mannheim Biochemicals, Indianapolis, Ind.), designed to detect pyuria and bacteriuria, was compared with culture of 1,020 unselected, consecutive midstream urine specimens and evaluated on its ability to predict colony counts at three levels. At the level of greater than or equal to 10(5) CFU/ml, the combined test (detection of leukocyte esterase and nitrite) had sensitivity of 82.3%, specificity of 67.9%, positive predictive value of 41.3%, and negative predictive value of 93.3% at prevalence rate of 21.6%. The test would have rejected 9.4% of the specimens with significant bacteriuria if the Chemstrip alone had been used.
Publisher
American Society for Microbiology
Reference13 articles.
1. Urine screening-to screen or not to screen;Bartlett R. C.;Clin. Microbiol. Newsl.,1984
2. Boehringer Mannheim Biochemicals. 1983. Urinalysis with CHEMSTRIP. Boehringer Mannheim Biochemicals Indianapolis Ind.
3. Galen R. S. and S. R. Gambino. 1975. Beyond normality: the predictive value and efficiency of medical diagnoses p. 115-116. John Wiley & Sons Inc. New York.
4. Culture of the urine;Hoeprich P. D.;J. Lab. Clin. Med.,1960
5. Pyelonephritis and bacteriuria. A major problem in preventive medicine;Kaas E. H.;Ann. Intern. Med.,1962