Affiliation:
1. Department of Pathology, Immunology and Laboratory Medicine, University of Florida
2. Shands Hospital at the University of Florida, Gainesville, Florida 32610
Abstract
ABSTRACT
We compared the FilmArray RP (Idaho Technology, Inc., Salt Lake City, UT) and the xTAG RVP (Luminex Corporation, Toronto, Canada) multiplex respiratory virus PCR methods for the detection of respiratory viruses in a set of 200 patient specimens frozen at −70°C after standard viral culture and antigen detection methods were done. Both systems detected between 40 to 50% more viruses than traditional methods, primarily rhinoviruses and human metapneumovirus. The FilmArray RP detected significantly more total viruses either alone or as part of mixed infections than the xTAG RVP, as well as an additional 21.6% more respiratory syncytial viruses. The xTAG RVP requires 5 to 6 h with 2.5 to 3 h of hands-on time, while the FilmArray RP takes about an hour with 3 to 5 min of hands-on time, making it much easier to perform.
Publisher
American Society for Microbiology
Cited by
120 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献