Affiliation:
1. Abbott Laboratories, Abbott Park, Illinois 60064
Abstract
ABSTRACT
Early detection of human immunodeficiency virus (HIV) in blood and blood products can be achieved by a sensitive nucleic acid amplification-based assay. We report on the performance of a PCR-based qualitative assay that detects both HIV type 1 (HIV-1) and HIV-2 with a sensitivity of 20 to 50 copies/ml. The assay has a specificity of 99.6% and an inhibition rate of 1.7%. One milliliter of sample is processed with a manifold system and Qiagen columns, and one-third of the extracted sample is used for PCR amplification. An internal control sequence, which is processed and amplified with each sample, monitors for amplification inhibition. Samples are reverse transcribed and are then amplified by reverse transcription-coupled PCR, after which HIV-1- and HIV-2-specific probes are hybridized to the amplified products. Following hybridization, samples are detected in the LCx instrument by microparticle enzyme immunoassay techniques. The detection system has an automated inactivation step that controls for PCR contamination. The HIV-1/2 qualitative RNA assay detects HIV-1 group M subtypes A, B, C, D, E, F, and G and group O. Testing of several HIV-1 seroconversion panels has demonstrated that the HIV-1/2 qualitative RNA assay detects HIV infection on the average of 6 days before p24 antigen can be detected and 11 days before antibodies can be detected.
Publisher
American Society for Microbiology
Cited by
21 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献