Affiliation:
1. Hyland-Immuno Division, Biomedical Research Center, Baxter, A-2304 Orth-Donau, Austria
Abstract
ABSTRACT
A real-time PCR method was developed to quantitate viral DNA that includes duplex amplification, internal standardization, and two-color fluorescence detection without the need to generate an external standardization curve. Applied to human parvovirus B19 DNA, the linear range was from 10
2
to at least 5 × 10
6
copies per ml of sample. The coefficient of variation was 0.29 using a run control of 2,876 copies per ml. The method reduces the risk of false-negative results, yields high precision, and is applicable for other DNA targets.
Publisher
American Society for Microbiology
Subject
Ecology,Applied Microbiology and Biotechnology,Food Science,Biotechnology
Cited by
49 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献