Affiliation:
1. Department of Pathology and Laboratory Medicine, Christ Hospital and Medical Center, Oak Lawn, Illinois 60453.
Abstract
Despite the customization of growth index thresholds within individual laboratories, use of the BACTEC NR 660 automated blood culture system results in a number of false-positive cultures. The results of Gram staining, acridine orange staining, and subculturing to agar media were evaluated on 210 false-positive blood cultures over a 6-month period. Inclusion of acridine orange staining in the routine workup of false-positive blood cultures can eliminate the need for subculturing.
Publisher
American Society for Microbiology
Reference2 articles.
1. Gornish N. and M. P. Weinstein. 1992. Comparison of Gram stain acridine orange stain and subculture in instrument-positive Bactec nonradiometric blood culture media C-88 p. 435. Abstr. 92nd Gen. Meet. Am. Soc. Microbiol. 1992. American Society for Microbiology Washington D.C.
2. Determination of growth value thresholds for BACTEC PLUS aerobic blood culture vials;McGowan J. E., Jr.;J. Clin. Microbiol.,1992
Cited by
6 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献