Affiliation:
1. Paul-Ehrlich-Institut, Paul-Ehrlich-Strasse 51-59, D-63225 Langen, Germany
Abstract
ABSTRACT
The performance of hepatitis E virus (HEV) RNA nucleic acid amplification (NAT)-based assays has been investigated using a panel of HEV-containing plasma samples. The panel comprised 22 HEV-positive plasma samples representing 10-fold serial dilutions of HEV genotypes 3a, 3b, 3f, and 4c obtained from blood donors. Two negative-control plasma samples were included. All samples were blinded. The plasma samples were prepared as liquid/frozen materials and distributed to participants on dry ice. Laboratories were requested to test the panel using their routine HEV assays and to score samples as either positive or negative and could optionally return data in copies/ml for HEV RNA. Twenty laboratories from 10 different countries participated in the study. Data were returned by all participating laboratories; 10 laboratories returned quantitative data. All assays except one were developed in-house using conventional or real-time reverse transcriptase PCR (RT-PCR) methodologies. There was a 100- to 1,000-fold difference in sensitivity between the majority of assays, independent of the virus strain. Although the quantitative data were limited, for the samples in the range of ∼6 to 4 log
10
copies/ml, the standard deviations of the geometric means of the samples ranged between 0.38 and 1.09. Except for one equivocal result, HEV RNA was not detected in the negative samples. The variability of assay sensitivity highlights the need for the standardization of HEV RNA NAT assays.
Publisher
American Society for Microbiology
Cited by
134 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献