Affiliation:
1. Division of Clinical Microbiology, Department of Laboratory Medicine and Pathology, Mayo Clinic, Rochester, Minnesota 55905
2. The University of Texas Health Science Center, Tyler, Texas 75708
Abstract
ABSTRACT
The SLOMYCO Sensititre panel and the custom JustOne strip (both from TREK Diagnostic Systems, Cleveland, OH) were evaluated for susceptibility testing of
Mycobacterium avium
complex isolates against clarithromycin. Seventy-one archived and prospectively collected isolates were tested using both the SLOMYCO panel and the JustOne strip, and the results were compared to those obtained using the BACTEC 460 (BD, Sparks, MD) radiometric method and a broth microdilution reference method. Results obtained by the SLOMYCO panel and the JustOne strip agreed with the BACTEC 460 method for 64/71 isolates (90%). Similarly, concordance with the broth microdilution method was 40/43 isolates (93%) for both test systems. The effect of the source medium on inoculum preparation was evaluated, and there were no differences noted in MICs, regardless of whether the inoculum was prepared from isolates grown in Middlebrook 7H9 medium, on Middlebrook 7H10 agar, or in VersaTREK broth culture bottles (Trek Diagnostics). Clarithromycin susceptibility testing of MAC using the SLOMYCO panel and the JustOne strip methods is easy to set up and simple to read and is readily incorporate into the clinical laboratory. These systems offer advantages over the BACTEC 460 system including the lack of a need for radioactive substrates, sharps, or costly instrumentation.
Publisher
American Society for Microbiology
Cited by
19 articles.
订阅此论文施引文献