Evaluation of six commercial kits for detection of human immunoglobulin M antibodies to Toxoplasma gondii. The FDA Toxoplasmosis Ad Hoc Working Group

Author:

Wilson M1,Remington J S1,Clavet C1,Varney G1,Press C1,Ware D1

Affiliation:

1. Division of Parasitic Diseases, National Center for Infectious Diseases, Centers for Disease Control and Prevention, Atlanta, Georgia 30341, USA. myw1@cdc.gov

Abstract

As a result of reports received by the Food and Drug Administration (FDA) of false-positive results obtained with FDA-cleared in vitro diagnostic kits for the detection of Toxoplasma-specific human immunoglobulin M (IgM) antibodies, an FDA-sponsored evaluation of six kits was performed. A battery of 258 serum specimens, including 30 specimens drawn 1 to 5 months after initial Toxoplasma infection and 228 specimens from Toxoplasma IgG-positive individuals, Toxoplasma IgG-negative individuals, rheumatoid factor-positive persons, and persons determined to be Toxoplasma IgM positive by commercially available assays, was assembled, randomly assorted, and coded. The battery was tested at the FDA with six commercially available kits, at the Palo Alto Medical Foundation (PAMF) by the PAMF double-sandwich IgM enzyme-linked immunosorbent assay (PAMF IgM ELISA), and at the Centers for Disease Control and Prevention (CDC) by the CDC EIA IgM. The results of the PAMF IgM ELISA that were obtained with the battery were considered to be the "gold standard" for this study; specificity rates were computed by considering the PAMF results to be 100% specific. Sensitivity and specificity rates were found to be as follows: CDC EIA IgM, 100 and 99.1%, respectively; Abbott IMx Toxo IgM, version 1, 100 and 77.5%, respectively; Abbott IMx Toxo IgM, version 2, 93.3 and 97.3%, respectively; Abbott Toxo-M EIA, 100 and 84.2%, respectively; BioMérieux Vitek VIDAS Toxo IgM, 100 and 98.6%, respectively; BioWhittaker Toxocap-M, 100 and 95.9%, respectively; Gull Toxo IgM, 97 and 85.6%, respectively; and Sanofi Diagnostics Pasteur Platelia Toxo IgM, 100 and 96.8%, respectively. Although the extent of false-positive reactions with these kits cannot be calculated because the study was retrospective and sample choices were biased, the results may be useful as an indicator of the relative specificities of these kits.

Publisher

American Society for Microbiology

Subject

Microbiology (medical)

Cited by 104 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3