Affiliation:
1. Roche Molecular Systems, Inc., Alameda, California 94501, USA.
Abstract
A panel of 136 genetically diverse group M and 5 group O adult isolates from outside the United States and Europe were evaluated by PCR with the Roche AMPLICOR HIV-1 test, a modified version of the AMPLICOR HIV-1 test, and a new primer pair/probe system. Detection of some of these isolates was less efficient with the AMPLICOR HIV-1 test; however, the assay was significantly improved by reducing the sample input and lowering the annealing temperature. The new primer pair/probe set detected 140 of 141 isolates, including the 5 group O isolates that were not detected with either of the AMPLICOR HIV-1 test formats.
Publisher
American Society for Microbiology
Cited by
22 articles.