Blood culture cross contamination associated with a radiometric analyzer

Author:

Griffin M R,Miller A D,Davis A C

Abstract

During a 9-day period in August 1980 in a New Jersey hospital, three pairs of consecutively numbered blood cultures from different patients were identified as positive for the same organism (two pairs of Klebsiella pneumoniae and one pair of group A Streptococcus), for each pair, both cultures were positive in the same atmosphere, both organisms had the same sensitivities, and the second of each pair grew at least 2 days after the first and was the only positive blood culture obtained from the patient. When the hospital laboratory discontinued use of its radiometric culture analyzer for 15 days, no more consecutive pairs of positive cultures occurred. Subsequent use of the machine for 9 days with a new power unit but the original circuit boards resulted in one more similar consecutive pair (Staphylococcus epidermidis). After replacement of the entire power unit, there were no further such pairs. Examination of the machine by the manufacturer revealed a defective circuit board which resulted in inadequate needle sterilization. Laboratories which utilize radiometric analyzers should be aware of the potential for cross contamination. Recognition of such events requires alert microbiologists and infection control practitioners and a record system in the bacteriology laboratory designed to identify such clusters.

Publisher

American Society for Microbiology

Subject

Microbiology (medical)

Reference7 articles.

1. Pseudobacteremia due to Staphylococcus aureus-New * York;Centers for Disease Control.;Morbid. Mortal. Weekly Rep.,1979

2. Mima polymorpha bacteremia: false-positive cultures due to contaminated penicillinase;Farls H. M.;J. Am. Med. Assoc.,1972

3. Klebsiella pneumoniae pseudobacteremia due to cross contamination of a radiometric blood culture analyzer;Greenhood G. P.;Infect. Control,1981

4. Pseudobacteremia caused by Clostridium sordellii;Lynch J. M.;Arch. Intern. Med.,1980

5. Evaluation of positive blood cultures. Guidelines for early differentiation of contamination from valid positive cultures;MacGregor R. R.;Arch. Intern. Med.,1972

Cited by 22 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. Bactec Blood Culture Systems;The Dark Art of Blood Cultures;2017-08-24

2. Skin Disease Presenting as an Outbreak of Pseudobacteremia in a Laboratory Worker;Journal of Clinical Microbiology;2001-01

3. Nosocomial Pseudo-outbreak ofFusarium verticillioidesAssociated With Sterile Plastic Containers;Infection Control & Hospital Epidemiology;2000-01

4. PSEUDOINFECTIONS;Infectious Diseases in Clinical Practice;1995-03

5. Pseudobacteraemia;Journal of Hospital Infection;1994-07