Doern G V,Herrmann J E,Henderson P,Stobbs-Walro D,Perron D M,Blacklow N R
Abstract
A total of 176 human fecal specimens were examined for the presence of rotavirus by four different assays: a monoclonal antibody enzyme immunoassay; the original polyclonal antibody enzyme immunoassay marketed by Abbott Laboratories, North Chicago, Ill. (Rotazyme I); a modification of this assay which is now commercially available (Rotazyme II); and a latex agglutination test (Rotalex) recently introduced by Medical Technology Corp., Somerset, N.J. In addition, selected specimens were examined for the presence of rotavirus by electron microscopy, immune electron microscopy, and RNA gel electrophoresis. A total of 40 specimens were positive in the monoclonal antibody enzyme immunoassay, and 136 were negative. Using the results obtained with this procedure as the reference standard, we found the sensitivities of the Rotazyme I, Rotazyme II, and Rotalex tests to be 97.4, 100, and 81.6%, respectively. The specificities of these three procedures were 88.8, 83.9, and 100%, respectively.
Publisher
American Society for Microbiology