A 1-year evaluation of Syva MicroTrak Chlamydia enzyme immunoassay with selective confirmation by direct fluorescent-antibody assay in a high-volume laboratory

Author:

Chan E L1,Brandt K1,Horsman G B1

Affiliation:

1. Laboratory and Disease Control Services, Saskatchewan Health, Regina, Canada.

Abstract

TThe Syva MicroTrak Chlamydia enzyme immunoassay (EIA; Syva Company, San Jose, Calif.) with cytospin and direct fluorescent-antibody assay (DFA) confirmation was evaluated on 43,630 urogenital specimens over a 1-year period in the Provincial Laboratory in Regina, Saskatchewan, Canada. This was a two-phase study intended to define a testing algorithm for Chlamydia trachomatis that would be both highly accurate and cost-effective in our high-volume (> 3,000 tests per month) laboratory. The prevalence of C. trachomatis infection in our population is moderate (8 to 9%). In phase 1, we tested 6,022 male and female urogenital specimens by EIA. All specimens with optical densities above the cutoff value and those within 30% below the cutoff value were retested by DFA. This was 648 specimens (10.8% of the total). A total of 100% (211 of 211) of the specimens with optical densities equal to or greater than 1.00 absorbance unit (AU) above the cutoff value, 98.2% (175 of 178) of the specimens with optical densities of between 0.500 and 0.999 AU above the cutoff value, and 83% (167 of 201) of the specimens with optical densities within 0.499 AU above the cutoff value were confirmed to be positive. A total of 12% (7 of 58) of the specimens with optical densities within 30% below the cutoff value were positive by DFA. In phase 2, we tested 37,608 specimens (32,495 from females; 5,113 from males) by EIA. Only those specimens with optical densities of between 0.499 AU above and 30% below the cutoff value required confirmation on the basis of data from phase 1 of the study. This was 4.5% of all specimens tested. This decrease in the proportion of specimens requiring confirmation provides a significant cost savings to the laboratory. The testing algorithm gives us a 1-day turnaround time to the final confirmed test results. The MicroTrak EIA performed very well in both phases of the study, with a sensitivity, specificity, positive predictive value, and negative predictive value of 96.1, 99.1, 90.3, and 99.7%, respectively, in phase 2. We suggest that for laboratories that use EIA for Chlamydia testing, a study such as this one will identify an appropriate optical density range for confirmatory testing for samples from that particular population.

Publisher

American Society for Microbiology

Subject

Microbiology (medical)

Reference5 articles.

1. Centers for Disease Control and Prevention. 1993. Recommendations for the prevention and management of Chlamydia trachomatis infections. Morbid. Mortal. Weekly Rep. 42(no. RR-12):1-39.

2. Direct fluorescentantibody confirmation of chlamydial antigen below the detection threshold of the Chlamydiazyme enzyme-linked immunosorbent assay;Kellog J. A.;J. Clin. Microbiol.,1993

3. Evaluation of Syva's enzyme immunoassay for the detection of Chlamydia trachomatis in urogenital specimens;Moncada J.;Diagn. Microbiol. Infect. Dis.,1992

4. Use of sequential enzyme immunoassay and direct fluorescent antibody tests for detection of Chlamydia trachomatis infections in women;Schwebke J. R.;J. Clin. Microbiol.,1990

5. Epidemiology of sexually transmitted Chlamydia trachomatis infections;Thompson S. E.;Epidemiol. Rev.,1983

Cited by 23 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3