Confirmation by 16S rRNA PCR of the COBAS AMPLICOR CT/NG Test for Diagnosis of Neisseria gonorrhoeae Infection in a Low-Prevalence Population

Author:

Diemert David J.1,Libman Michael D.1,Lebel Pierre1

Affiliation:

1. Department of Microbiology, Montreal General Hospital, McGill University Health Centre, Montreal, Canada

Abstract

ABSTRACT The COBAS AMPLICOR CT/NG test is widely used for the diagnosis of Neisseria gonorrhoeae infection using genital swabs or urine samples. Although highly specific, cross-reactivity occurs with some nonpathogenic strains of Neisseria and Lactobacillus species. In low-prevalence populations, even highly specific assays may require confirmatory testing of positive results. We assessed the positive predictive value (PPV) of this test in a low-prevalence (0.5%) setting. Genital and urine specimens testing positive using the COBAS AMPLICOR NG test were retested using an investigational 16S rRNA PCR assay. Additionally, 737 specimens were tested in parallel by both culture and the above PCR protocol. Of 9,772 specimens tested in-house, 168 were positive by the AMPLICOR test; in addition, 62 AMPLICOR-positive specimens were referred to our laboratory for confirmatory testing, yielding 230 positive specimens. Of these, 72 were confirmed positive by 16S rRNA PCR, yielding a specificity of 98.7% and a PPV of 31.3%. Specificity was similar for all specimen types, whereas PPV varied with prevalence: specimens from males, females, urine specimens, and genital swabs had PPVs of 70.8, 13.3, 51.9, and 20.1%, respectively. The PPV was higher when the initial AMPLICOR optical density (OD) was ≥3.5 versus initial and repeat OD readings in an equivocal zone of ≥0.2 to <3.5 (65.1 versus 10.1%; P < 0.001). On repeat testing of specimens with ODs in the equivocal zone, 54 gave ODs of ≥0.2 and <2.0, 35 gave ODs of ≥2.0 and <3.5, and 12 gave ODs of ≥3.5, with 3.7, 20, and 33.3% confirmed positive, respectively ( P = 0.004). Comparing PCR to culture as the “gold standard,” specificity increased from 96.8 to 99.9% when 16S rRNA PCR was performed on specimens positive by the COBAS AMPLICOR NG test. Confirmatory testing with a more specific method such as 16S rRNA PCR should be considered in low-prevalence populations, especially for specimens with an OD in the equivocal zone.

Publisher

American Society for Microbiology

Subject

Microbiology (medical)

Cited by 50 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3