Continued Stabilization of a Cementless 3D-Printed Total Knee Arthroplasty

Author:

van der Lelij Thies J.N.1ORCID,Marang-van de Mheen Perla J.2ORCID,Kaptein Bart L.1ORCID,Toksvig-Larsen Sören34ORCID,Nelissen Rob G.H.H.1ORCID

Affiliation:

1. Department of Orthopaedics, Leiden University Medical Center, Leiden, The Netherlands

2. Medical Decision Making, Department of Biomedical Data Sciences, Leiden University Medical Center, Leiden, The Netherlands

3. Department of Orthopaedics, Hässleholm Hospital, Hässleholm, Sweden

4. Department of Clinical Sciences, Lund University, Lund, Sweden

Abstract

Background: Three-dimensional (3D) printing of highly porous orthopaedic implants aims to promote better osseointegration, thus preventing aseptic loosening. However, short-term radiostereometric analysis (RSA) after total knee arthroplasty (TKA) has shown higher initial migration of cementless 3D-printed tibial components compared with their cemented counterparts. Therefore, critical evaluation of longer-term tibial component migration is needed. We investigated migration of a cementless 3D-printed and a cemented tibial component with otherwise similar TKA design during 5 years of follow-up, particularly the progression in migration beyond 2 years postoperatively. Methods: Seventy-two patients were randomized to a cementless 3D-printed Triathlon Tritanium (Stryker) cruciate-retaining (CR) TKA or a cemented Triathlon CR (Stryker) TKA implant. Implant migration was evaluated with RSA at baseline and postoperatively at 3 months and at 1, 2, and 5 years. The maximum total point motion (MTPM) of the tibial component was compared between the groups at 5 years, and progression in migration was assessed between 2 and 5 years. Individual implants were classified as continuously migrating if the MTPM was ≥0.1 mm/year beyond 2 years postoperatively. Clinical scores were evaluated, and a linear mixed-effects model was used to analyze repeated measurements. Results: At 5 years, the mean MTPM was 0.66 mm (95% confidence interval [CI], 0.56 to 0.78 mm) for the cementless group and 0.53 mm (95% CI, 0.43 to 0.64 mm) for the cemented group (p = 0.09). Between 2 and 5 years, there was no progression in mean MTPM for the cementless group (0.02 mm; 95% CI, −0.06 to 0.10 mm) versus 0.07 mm (95% CI, 0.00 to 0.14) for the cemented group. One implant was continuously migrating in the cementless group, and 4 were continuously migrating in the cemented group. The clinical scores were comparable between the groups across the entire time of follow-up. Conclusions: No significant difference in mean migration was found at 5 years between the cementless and cemented TKA implants. Progression of tibial component migration was present beyond 2 years for the cemented implant, whereas the cementless implant remained stable after initial early migration. Level of evidence: Therapeutic Level I. See Instructions for Authors for a complete description of levels of evidence.

Publisher

Ovid Technologies (Wolters Kluwer Health)

Subject

Orthopedics and Sports Medicine,General Medicine,Surgery

Reference40 articles.

1. Increased initial cement-bone interlock correlates with reduced total knee arthroplasty micro-motion following in vivo service;Miller;J Biomech.,2014

2. Implant Interface Debonding After Total Knee Arthroplasty: A New Cause for Concern?;Sadauskas;Arthroplast Today.,2020

3. Cementless total knee arthroplasty: current concepts review;Dalury;Bone Joint J.,2016

4. Long-term implant survivorship of cementless total knee arthroplasty: a systematic review of the literature and meta-analysis;Mont;J Knee Surg.,2014

5. Topological design and additive manufacturing of porous metals for bone scaffolds and orthopaedic implants: A review;Wang;Biomaterials.,2016