Affiliation:
1. Department of Orthopaedic Surgery, Washington University in St. Louis, St. Louis, Missouri
2. Department of Orthopaedic Surgery, University of Arkansas for Medical Sciences, Little Rock, Arkansas
3. Orthopaedic Associates of Hartford, Hartford, Connecticut
4. Department of Orthopaedic Surgery, Northwestern University, Chicago, Illinois
Abstract
Background:
Total elbow arthroplasty (TEA) remains a valuable tool for treating inflammatory, degenerative, and traumatic elbow conditions. This study aimed to understand the incidence of and risk factors for reoperation following TEA at a high-volume center utilizing an implant with a convertible linkage and the potential for anatomic lateral column reconstruction.
Methods:
All patients undergoing primary TEA with the Latitude prosthesis (Stryker) from July 2001 to May 2020 were identified. Patient characteristics, the surgical indication, and implant characteristics were obtained. Additionally, the indications and timing were identified for reoperations. Postoperative radiographs were analyzed to assess cement quality, implant position, lateral column reconstruction, and distal humeral bone loss.
Results:
Two hundred and nineteen TEAs were performed in 206 patients; 162 (74.0%) were in female patients, the mean patient age was 61 years (range, 23 to 95 years), and the mean follow-up was 11.8 years (range, 2 to 21 years). The most common indication for surgery was sequelae of trauma (36%). Ninety elbows (41.1%) required 200 reoperations at a mean of 19.6 months. Forty TEAs (18.3%) required revision of nonmodular implants, with 14 (6.4%) undergoing definitive explantation. The 5 and 10-year survivorship was 86.1% and 79.7%, respectively. The most common of the reasons for revision was aseptic loosening (53.8%), and the most common cause for non-revision reoperation was infection (23%). Younger age and greater follow-up duration were associated with greater revision and reoperation risks. Aseptic ulnar loosening was associated with a short ulnar stem length and component linkage (30% of short linked ulnar stems loosened; p < 0.001). Aseptic humeral loosening was associated with less-than-adequate cement-mantle quality (p = 0.04).
Conclusions:
Extended follow-up of the Latitude prosthesis at a high-volume center demonstrates that TEA continues to be hampered by a high reoperation rate, primarily due to infection and aseptic loosening. Technical factors such as good cement-mantle quality, longer stem length, and unlinked implants may play a role in preventing aseptic loosening. Further work is required to optimize long-term outcomes following TEA through improved understanding of appropriate surgical indications, techniques, and implant utilization.
Level of Evidence:
Therapeutic Level IV. See Instructions for Authors for a complete description of levels of evidence.
Publisher
Ovid Technologies (Wolters Kluwer Health)