Author:
Attar Tarik,Harek Yahia,Larabi Lahcene
Abstract
 A selective and sensitive method for determination of copper in blood by adsorptive differential pulse cathodic stripping voltammetry is presented. The method is based on adsorptive accumulation of the complexes of Cu (II) ions with benzenesulfonyl hydrazide onto hanging mercury drop electrode (HMDE), followed by the reduction of the adsorbed species by differential pulse cathodic stripping voltammetry. The effect of various parameters such as supporting electrolyte, concentration of benzenesulfonyl hydrazide, accumulation potential, accumulation time and stirring rate on the selectivity and sensitivity were studied. The optimum conditions for determination of copper include perchloric acid 0.03 M, concentration of benzenesulfonyl hydrazide 7.5×10-5 M, the accumulation potential of -350 mV (vs. Ag/AgCl), the accumulation time of 50 s, and the scan rate of 50 mV s-1. Under optimized conditions, linear calibration curves were established for the concentration of Cu (II) in the range of 0.62-275 ng mL-1, with detection limit of 0.186 ng mL-1 for Cu (II). The procedure was successfully applied to the determination of copper ion in whole blood samples.
Publisher
Mediterranean Journal of Chemistry
Cited by
9 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献