Refocus flip angle modulation on the pd tse sequences in the magnetic resonance imaging of the knee, for the evaluation of meniscal injuries

Author:

Curatolo Calogero1ORCID,Lisanti Sara2,Daricello Marco3,Caruso Virginia3,Candela Fabrizio4,Cimino Pietro4,Spoto Italia4,Cutaia Giuseppe5ORCID,Lo Re Giuseppe6ORCID,Galia Massimo7ORCID

Affiliation:

1. Dipartimento di Diagnostica per immagini, A.O.U.P. "P. Giaccone", Palermo

2. Sezione di Scienze Radiologiche, BIND, Università degli Studi di Palermo, Italia

3. Dipartment of Radiology, A.O.U.P. "P. Giaccone", Palermo

4. Dipartimento di Diagnostica per Immagini, A.O Universitaria Policlinico “Paolo Giaccone”, Palermo, Italia

5. Sezione di Scienze Radiologiche, BIND, Università degli Studi di Palermo, Italia; Section of Gastroenterology & Hepatology, Department of Healh Promotion, Mother and Child

6. Radiology Section, BIND, University of Palermo

7. Sezione di Scienze radiologiche, BIND, Università degli Studi di Palermo

Abstract

The Radiofrequency Refocused Echo Spin-Echo Echo-Train Rapid-Acquisition sequences, known as Turbo Spin Echo or Fast Spin Echo are the most commonly used sequences in Magnetic Resonance as they provide a notable contribution in morphological and anatomical terms, thanks to their high spatial and contrast resolution. Furthermore, they allow a higher signal noise ratio than the other families of sequences thanks to the numerous refocusing pulses and thanks to a TR so long as to allow sampling as many K-space phase encodings as possible and with a complete recovery of the Longitudinal Magnetization. Finally, the multiple 180 ° pulses of these sequences reduce the inhomogeneities of the Magnetic Field by minimizing the phenomena of magnetic susceptibility. However, their application, especially in scanners with a high intensity of static magnetic field B0(1.5T and/or 3T)is prevented by the deposition of RF due to the long echo trains, which sometimes involve exceeding the limits of the specific absorption rate for patient safety. Over time, a common solution to the SAR problem has been the use of refocusing angles smaller than 180°(160°-140°-120°),which lead to its reduction at the cost, however, of an obvious penalty in terms of signal-to-noise ratio. In this study we present a modulation method of the Refocus Flip-Angle applied to the DP-TSE sequences in the evaluation of meniscal lesions in the study protocol of the Magnetic Resonance of the Knee, which exploits the phenomenon of the Pseudo-Steady-State(PSS), leading to a noticeable SARreduction without loss of SNR and also providing excellent contrast resolution

Publisher

Ordine TSRM PSTRP di Napoli, Avellino, Benevento

Subject

Earth-Surface Processes

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3