Validation of a novel automated system, Fluispotter®, for serial sampling of dried blood spots

Author:

Krogh Jesper12,Plomgaard Peter23,Frikke-Schmidt Ruth23,Velschow Sten4,Johannesen Jesper25,Hilsted Linda Maria3,Schrøder Malene4,Feldt-Rasmussen Ulla12ORCID

Affiliation:

1. Department of Endocrinology & Metabolism, Copenhagen University Hospital - Rigshospitalet, Copenhagen, Denmark

2. Department of Clinical Medicine, University of Copenhagen, Copenhagen, Denmark

3. Department of Clinical Biochemistry, Copenhagen University Hospital - Rigshospitalet, Copenhagen, Denmark

4. Fluisense ApS, Lillerød, Denmark

5. Department of Pediatrics, Copenhagen University Hospital - Herlev & Gentofte, Copenhagen, Denmark

Abstract

Repeated blood sampling is required in certain clinical and research settings, which is currently performed by drawing blood from venous catheters requiring manual handling of each sample at the time of collection. A novel body-worn device for repeated serial samples, Fluispotter®, with automated extraction, collection, and storage of up to 20 venous dried blood spot samples over the course of 20 h may overcome problems with current methods for serial sampling. The purpose of this study was to assess the performance and safety of Fluispotter for the first time in healthy subjects. Fluispotter consists of a cartridge with tubing, a reservoir for flushing solution, pumps and filterpaper, and a multi-lumen catheter placed in the brachial vein. We recruited healthy subjects for testing in an in-hospital setting. Fluispotter was attached by an anesthesiologist to 22 healthy subjects of which 9/22 (40.9%) participants had all 20 samples taken, which was lower than the goal of complete sampling in 80% of the subjects (P = 0.02). The main reason for sample failure was clogging of blood flow which was observed in 11/22 (50%) of the participants. No serious adverse events occurred, and the participants rated the pain from the insertion and the removal of catheter as very low. A cortisol profile showed nadir values at midnight and highest values at 05:00 h. Although full sampling was not successful in all participants, the Fluispotter technology proved safe and highly acceptable to the participants producing the expected cortisol profile without the requirement of staff during sample collection.

Publisher

Bioscientifica

Subject

Endocrinology,Endocrinology, Diabetes and Metabolism,Internal Medicine

Reference22 articles.

1. Diagnosis and treatment of primary adrenal insufficiency: an Endocrine Society clinical practice guideline;Bornstein,2016

2. Acromegaly: an endocrine society clinical practice guideline;Katznelson,2014

3. Evaluation and treatment of adult growth hormone deficiency: an Endocrine Society clinical practice guideline;Molitch,2011

4. Advanced circadian phase in mania and delayed circadian phase in mixed mania and depression returned to normal after treatment of bipolar disorder;Moon,2016

5. Circadian disruption: new clinical perspective of disease pathology and basis for chronotherapeutic intervention;Smolensky,2016

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3