Abstract
A small-angle neutron scattering (SANS) instrument at the China Spallation Neutron Source (CSNS) is an operating instrument for studying structures and inhomogeneities with dimensions ranging from 1 to 100 nm. Preparing multiple samples at once and measuring them sequentially is a common approach in SANS experiments to reduce neutron beamline wastes and increase experimental efficiency. We present the development of an automatic sample changer for the SANS instrument, including system design, thermal simulation, optimization analysis, structure design details, and temperature control test results. It features a two-row construction that can hold 18 samples on each row. The controllable temperature range is −30 to 300 °C. Furthermore, neutron scattering experiments on SANS at CSNS proved that this instrument has good temperature control performance and low background. This automatic sample changer is optimized for usage at SANS and will be offered to other researchers through the user program.
Funder
National Key Research and Development Program of China
Youth Innovation Promotion Association
National Natural Science Foundation of China
the Ninth Batch of Innovation and Entrepreneurship Leading Talentsin 2019
Guangdong Natural Science Funds for Distinguished Young Scholar
Guangdong Provincial Key Laboratory of Extreme Conditions
Guangdong Basic and Applied Basic Research Foundation
Cited by
3 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献