Wehnelt photoemission in an ultrafast electron microscope: Stability and usability

Author:

Willis Simon A.12ORCID,Curtis Wyatt A.12,Flannigan David J.12ORCID

Affiliation:

1. Department of Chemical Engineering and Materials Science, University of Minnesota 1 , 421 Washington Avenue SE, Minneapolis, Minnesota 55455, USA

2. Minnesota Institute for Ultrafast Science, University of Minnesota 2 , Minneapolis, Minnesota 55455, USA

Abstract

We tested and compared the stability and usability of three different cathode materials and configurations in a thermionic-based ultrafast electron microscope: (1) on-axis thermionic and photoemission from a custom 100 μm diameter LaB6 source with a graphite guard ring, (2) off-axis photoemission from the Ni aperture surface of the Wehnelt electrode, and (3) on-axis thermionic and photoemission from a custom 200 μm diameter polycrystalline Ta source. For each cathode type and configuration, including the Ni Wehnelt aperture, we illustrate how the photoelectron beam-current stability is deleteriously impacted by simultaneous cooling of the source following thermionic heating. Furthermore, we demonstrate usability via collection of parallel- and convergent-beam electron diffraction patterns and by formation of the optimum probe size. We find that usability of the off-axis Ni Wehnelt-aperture photoemission is at least comparable to on-axis LaB6 thermionic emission, as well as to on-axis photoemission [the heretofore conventional approach to ultrafast electron microscopy (UEM) in thermionic-based instruments]. However, the stability and achievable beam currents for off-axis photoemission from the Wehnelt aperture were superior to that of the other cathode types and configurations, regardless of the electron-emission mechanism. Beam-current stability for this configuration was found to be ±1% (one standard deviation from the mean) for 70 min (longest duration tested), and steady-state beam current was reached within the sampling-time resolution used here (∼1 s) for 15 pA beam currents (i.e., 460 electrons per packet for a 200 kHz repetition rate). Repeatability and robustness of the steady-state condition were also found to be within ±1% of the mean. We discuss the implications of these findings for UEM imaging and diffraction experiments, for pulsed-beam damage measurements, and for practical switching between optimum conventional TEM and UEM operation within the same instrument.

Funder

U.S. Department of Energy

National Science Foundation

Publisher

AIP Publishing

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3