Affiliation:
1. State Key Laboratory of Electric Insulation and Power Equipment, Xi’an Jiaotong University, Xi’an 710049, People’s Republic of China
Abstract
Resonance frequency drift caused by a change in temperature greatly limits the application of high- Q resonators with high temperature sensitivity in photoacoustic (PA) gas detection systems. In this work, a chirp-wavelength combined modulation method was designed by incorporating a real-time frequency scanning in wavelength-modulated PA spectroscopy to reduce the influence of temperature changes on measurement. Theoretical analysis shows that the chirp rate depends on the precision requirements and the cutoff frequency of the cascaded low-pass filter. Trace acetylene measurement experiment at varying temperature verified that the proposed method can significantly reduce the temperature sensitivity within a preset temperature range. Thus, this method can effectively reduce the temperature sensitivity of a high- Q resonator for improving the measurement accuracy and detection limit in trace gas detection.
Funder
Science and Technology Project of State Grid
Cited by
4 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献