Affiliation:
1. National Institute of Standards and Technology, Physical Measurement Laboratory, Applied Physics Division, Boulder, CO 80305, USA
Abstract
To calibrate laser power and energy meters, the National Institute of Standards and Technology (NIST) uses several detector-based realizations of the scale for optical radiant flux; these realizations are appropriate for specific laser power/energy ranges and optical coupling configurations. Calibrations from 1 μW to 2 W are currently based upon calorimeters. Validation by comparisons against other primary representations of the optical watt over the last two decades suggests the instruments operate well within their typical reported uncertainty level of 0.86 % with 95 % confidence. The dominant uncertainty contribution in the instrument is attributable to light scattered by the legacy window, which was not previously recognized. The inherent electro-optical inequivalence in the calorimeter’s response was reassessed by thermal modeling to be 0.03 %. The principal contributions to the overall inequivalence were corrected, yielding a shift in scale representation under 0.2 % for typical calibrations. With updates in several uncertainty contributions resulting from this reassessment, the resulting combined expanded uncertainty (k = 2) is 0.84 %, which is essentially unchanged from the previous result provided to calibration customers.
Funder
Physical Measurement Laboratory
Publisher
National Institute of Standards and Technology (NIST)
Cited by
1 articles.