Power-injectable totally implantable venous access devices – analysis of success and complication rates of ultrasound-guided implantation and a patient satisfaction survey

Author:

Kunz-Virk Julia1,Krüger Karsten1

Affiliation:

1. Department of Radiology and Interventional Therapy, Vivantes Humboldt Hospital and Spandau Hospital, Berlin, Germany

Abstract

Summary. Background: To retrospectively evaluate the success rates, peri-interventional, early and late complications and patient satisfaction associated with power-injectable totally implantable venous access devices (TIVAPs). Patients and methods: Between April 2011 and March 2016, a total of 1,203 TIVAPs were implanted in 1,169 patients. Ultrasound-guided, fluoroscopically controlled implantation was performed through the subclavian or internal jugular vein. The systematic analysis focused on the rate of successful port implantations, the frequency of peri-interventional, early and late complications and on how the experience of the implanting radiologist impacts these parameters. Additionally, a standardized questionnaire was administered to the 102 study patients in a telephone interview to survey their subjective rating of the port implantation. Results: 99.5 % of TIVAPs were implanted successfully. In 4 out of 6 patients, the implantation was repeated successfully at a later time. Complication rates were 1.4 % (0.0512/1.000 catheter days) for peri-interventional, 2.9 % (0.081 per 1,000 catheter days) for early and 8.3 % (0.2288 per 1,000 catheter days) for late complications. The radiologist’s experience level and vein selection did not have a significant impact. Most peri-interventional complications (82.4 %) were of minor severity. The early (61.5 %) and late (65.6 %) complications were more frequently of major severity. Interventions to manage complications comprised port explanation in 46.9 %, conservative therapy in 17.4 % and interventional therapy in 12.2 %. At 1 and 3 months after port placement, the majority were satisfied or very satisfied with the interventional port implantation. Conclusions: Ultrasound-guided, fluoroscopically controlled implantation of TIVAPs is a safe procedure with low complication rates, high success rates and high patient satisfaction.

Publisher

Hogrefe Publishing Group

Subject

Cardiology and Cardiovascular Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3